ResMed S8 AutoSet Spirit User Manual page 39

Table of Contents

Advertisement

Problem
Possible Cause
Display error message:
The air tubing is loose or
blocked.
Check tube!!
Key if done
Display error message:
Component failure.
SYSTEM ERROR
Call service!
Display error message:
You have experienced
excessively high leak levels
High leak in
during the night.
last session
Display error message:
You have experienced
excessively high leak levels for
HIGH LEAK!!!
more than 20 seconds.
Adjust Mask
Display error message:
Component failure.
Exxxx (where xxxx
defines an error)
The following message
The Data Card is not inserted
is displayed on the
correctly.
LCD after you try to
update settings using
the Data Card:
Card Error
Remove Card
The Data Card is only partly
inserted.
Solution
Check that the air tubing is
connected securely to your mask
and the air outlet on the front of
the S8 AutoSet Spirit.
When you have checked the air
tubing, press the Start/Stop key to
restart the device.
If this does not clear the message,
disconnect the power cord and
then reconnect it to restart the
device.
Return your S8 AutoSet Spirit for
servicing.
Check that your air tubing is
connected properly and that your
mask does not leak excessively.
Use the mask-fitting feature to
help you to fit your mask
properly.
If this message appears again,
contact your clinician.
Check that your air tubing is
connected properly and that your
mask does not leak excessively.
Use the mask-fitting feature to
help you to fit your mask
properly.
If this message appears again,
contact your clinician.
Return your S8 AutoSet Spirit for
servicing.
Ensure that the Data Card is
inserted with the arrow facing up
and pointing towards the device.
Ensure that the device is switched
on and in standby mode before
you insert the Data Card.
Ensure that the Data Card is
inserted into the slot on the Data
Card module as far as it will go.
T
ROUBLESHOOTING
33