System Reference Manual
7P00370-H01
NOTICE
If you intend to connect the RJ-45 port to a router or switch, ensure the GC's DHCP
server is turned off by flipping the SW1 switch to OFF. Connecting to your LAN with
the DHCP server on will disrupt your local network's functioning.
NOTICE
The GC can be connected (or remain connected) to a local network via the wired
Ethernet 2 port on the back plane while the DHCP feature is being used.
2.
3.
NOTICE
If your computer is not configured for dynamic IP addressing (static IP), you will
need to configure the 370XA with a static IP address that is in the same subnet as
your computer. Contact your network administrator for these IP settings and enter
the required settings into the TCP/IP Settings screen on the 370XA.
4.
NOTICE
You must have at least MON2020 version 3.0 to communicate to the 370XA.
5.
6.
7.
3 - 52
Connect between the GC's RJ-45 connector and the local
computer with a standard Cat 5 Ethernet cable.
Wait 30 seconds for the computer to update its TCP/IP
settings.
Start MON2020.
Select Connect on the Chromatograph menu. The Connect
to GC window opens.
Click the Ethernet button in the Direct-DHCP row. The Login
screen opens.
Enter your login information and press OK. The computer will
connect to the GC. The status bar at the bottom of the
MON2020 screen will show the GC name, the alarm status, the
mode and the time and date of the connected GC.
370XA Gas Chromatograph
APRIL 2014