370XA Gas Chromatograph
APRIL 2014
While the GC is heating up to temperature, the LOI will show the
Calibration Gas Info composition screen so that you can confirm that
the module's calibration gas values match the concentrations and the
uncertainty on the calibration gas bottle certificate.
When the oven temperature has stabilized, the software assistant will
automatically run calibration gas and validate the module. The
validation will run three analysis cycles of calibration gas and confirm
that the analysis is within the pre-configured specifications. If the
analysis is within the specifications, a calibration cycle will then be
System Reference Manual
7P00370-H01
4 - 43