370XA Gas Chromatograph
APRIL 2014
NOTICE
The washer is slightly curved and should be placed so that it curves up to the center.
29.
30.
31.
32.
33.
Use a torque wrench and 11mm socket to tighten the assembly
bolt to 20 ft/lb (2.76 kg/m).
Set the new o-rings into the actuation valve ports on the
module base plate.
Insert the three socket screws into the holes of the analytical
valve base plate.
Align the thick and thin locating pins in the base plate with the
locating holes in the analysis valve and install the analysis
valve onto the base plate.
Tighten the socket screws with the 2.5 mm hex or allen
wrench.
System Reference Manual
7P00370-H01
6 - 21