System Reference Manual
7P00370-H01
5.
6.
7.
8.
9.
10.
6 - 4
Remove the solenoid and the seal.
Inspect the new solenoid's seal and ensure that it is firmly
seated in the base of the solenoid.
Place the solenoid onto the module base with the connector
located closest to the module edge.
Hand-tighten the two hex screws to secure the solenoid.
Reconnect the connector to the solenoid, ensuring the
retention clip engages. The connectors are labeled according
to the following table:
Label
SV1
SV2
SV3
SV 4
Reinstall the module on the 370XA base and analyze gas
through each stream to confirm the repair has been successful.
See
"After you make the repairs, you must test the module and
adjust the analytical method (valve timings, integration events,
and calibration settings). See "Auto valve timing" on page 4-
37." on page
6-50.
370XA Gas Chromatograph
Solenoid
Calibration solenoid
Stream 1
Stream 2
Stream 3
APRIL 2014