To trigger on a specific field of the video signal
To examine the components of a video signal, trigger on either Field 1 or Field
2. When a specific field is selected, the oscilloscope triggers on the rising edge
of the first serration pulse in the vertical sync interval in the specified field (1
or 2).
1 Press the Trigger
More
2 Press the
softkey, then press the
Settings
appropriate TV standard.
3 Press the
Mode
softkey and select
Triggering on Field 1
key, then press the
TV
Standard
Field1
or
Field2
Triggering the Oscilloscope
To use TV triggering
softkey.
softkey to select the
.
3- 35