Agilent Technologies 54622A User Manual page 218

Oscilloscopes and mixed-signal oscilloscopes
Hide thumbs Also See for 54622A:
Table of Contents

Advertisement

* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and
±10 °C from firmware calibration temperature.
Vertical System: Analog Channels
54621A/D, 54622A/D
54624A
54621A/D
Bandwidth (-3dB)*
ac coupled
Calculated rise time
54622A/D, 54624A
Bandwidth (-3dB)*
ac coupled
Calculated rise time
Single Shot Bandwidth
1
Range
Maximum Input
Offset Range
Dynamic Range
Input Resistance
Input Capacitance
Coupling
BW Limit
Channel-to-Channel Isolation
Probes
Probe ID (Agilent/HP &
Tek Compatible)
1
1 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting.
Performance Characteristics
Vertical System: Analog Channels
Ch1 and 2 simultaneous acquisition
Ch 1, 2, 3 and 4 simultaneous acquisition
dc to 60 MHz
3.5 Hz to 60 MHz
~5.8 ns (= 0.35/bandwidth)
dc to 100 MHz
3.5 Hz to 100 MHz
~3.5 ns (= 0.35/bandwidth)
50 MHz
1 mV/div to 5 V/div
CAT I 300 Vrms, 400 Vpk
CAT II 100 Vrms, 400 Vpk
with 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
±5 V on ranges <10 mV/div
±25 V on ranges 10 mV/div to 199 mV/div
±100 V on ranges
Lesser of ±8 div or ±32 V
1 M
±1%
~ 14 pF
ac, dc, ground
~ 20 MHz selectable
dc to 20 MHz > 40 dB (with channels at same V/div)
20 MHz to max bandwidth > 30 dB
10:1 10074C shipped standard for each scope channel
Auto probe sense
7- 4
200 mV/div