Agilent Technologies 54622A User Manual page 91

Oscilloscopes and mixed-signal oscilloscopes
Hide thumbs Also See for 54622A:
Table of Contents

Advertisement

6 If you select a line mode when using standards NTSC, PAL, PAL-M, or
SECAM, press the
the line number on which you want to trigger.
7 If you select a line mode when using the Generic standard, press the
softkey, then rotate the Entry knob to select the desired count
Count #
number.
Listed below are the line (or count) numbers per field for each TV standard.
Line (or count for Generic) numbers per field for each TV standard
TV standard
NTSC
PAL
PAL-M
SECAM
Generic
softkey, then rotate the Entry knob to select
Line #
Field 1
Field 2
1 to 263
1 to 262
1 to 313
314 to 625
1 to 263
264 to 525
1 to 313
314 to 625
1 to 1024
1 to 1024
Triggering the Oscilloscope
To use TV triggering
Alt Field
1 to 262
1 to 312
1 to 262
1 to 312
1 to 1024 (vertical)
3- 31