Overshoot
Overshoot is distortion that follows a major edge transition expressed as a
percentage of Amplitude. The X cursors show which edge is being measured
(edge closest to the trigger point).
Rising edge overshoot =
Falling edge overshoot =
Overshoot
Automatic Overshoot measurement
Making overshoot and preshoot measurements
local Maximum – Top
----------------------------------------------------- X 100
Amplitude
Base – local Minimum
------------------------------------------------------ X 100
Amplitude
local Maximum
local Minimum
Overshoot
Making Measurements
Top
Base