* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and
±10 °C from firmware calibration temperature.
Analog Channel Triggering
Range (Internal)
Sensitivity*
Coupling
Logic (D15 - D0) Channel Triggering (54621D and 54622D)
Threshold Range (used defined)
Threshold Accuracy*
Predefined Thresholds
External (EXT) Triggering
Input Resistance
Input Impedance
Maximum Input
Range
Sensitivity
Coupling
Performance Characteristics
Analog Channel Triggering
±6 div
Greater of 0.35 div or 2.5 mV
ac (~3.5 Hz), dc, noise reject, HF reject and LF reject (~ 50 kHz)
±8.0 V in 10 mV increments
±(100 mV + 3% of threshold setting)
TTL = 1.4 V, CMOS = 2.5 V, ECL = -1.3 V
1 M
, ±3%
~ 14 pF
CAT I 300 Vrms, 400 Vpk
CAT II 100 Vrms, 400 Vpk
with 10074C 10:1 probe:CAT I 500 Vpk, CAT II 400 Vpk
±10 V
dc to 25 MHz, < 50 mV
25 MHz to max bandwidth, < 150 mV
ac (~ 3.5 Hz), dc, noise reject, HF reject and LF reject (~ 50 kHz)
7- 8