* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and
±10 °C from firmware calibration temperature.
General Characteristics
Physical:
Size
Weight
Calibrator Output
Trigger Out
Printer Power
Power Requirements
Line Voltage Range
Line Frequency
Power Usage
Environmental Characteristics
Ambient Temperature
Humidity
Altitude
Vibration
Shock
Pollution degree2
Indoor use only
Installation categories
32.26 cm wide x 17.27 cm high x 31.75 cm deep (without handle)
~ 6.35 kgs (14 lbs)
Frequency ~1.2 kHz; Amplitude 5 V
Ω
0 to 5 V with 50
source impedance; delay ~ 55 ns
7.2 to 9.2 V, 1 A
100 - 240 VAC ±10%, CAT II, automatic selection
47 to 440 Hz
100 W max
Operating -10 °C to +55 °C
Non-operating -51 °C to +71 °C
Operating 95% RH at 40 °C for 24 hr
Non-operating 90% RH at 65 °C for 24 hr
Operating to 4,570 m (15,000 ft)
Non-operating to 15,244 m (50,000 ft)
HP/Agilent class B1 and MIL-PRF-28800F Class 3 random
HP/Agilent class B1 and MIL-PRF-28800F (operating 30 g, 1/2 sine, 11-ms duration, 3
shocks/axis along major axis. Total of 18 shocks)
Normally only dry non-conductive pollution occurs. Ocassionally a temporary
conductivity caused by condensation must be expected.
This instrument is rated for indoor use only
CAT I: Mains isolated
CAT II: Line voltage in appliance and to wall outlet
Performance Characteristics
General Characteristics
7- 11