HELIX 9600 Installation, Operation And Maintenance Instructions page 53

High voltage compressor
Table of Contents

Advertisement

©2007 Helix Technology Corporation
5. Move the adsorber from under the mounting tabs in the base as
shown in
Figure 5-2
6. Install the replacement adsorber under the mounting tabs and
secure it into place with the bolt removed during Step 4.
7. Using two wrenches as shown in
sealing couplings quickly to minimize helium leakage.
8. Install the Compressor rear panel.
9. Ensure that the pressure gauge reads the proper value as shown in
Table
4-1. If additional gas pressure is required, refer to
Helium"
in this section. If gas pressure needs to be reduced, refer to
"Reducing Helium Pressure"
10. Record the adsorber replacement date on the label as shown in
Figure
5-2, and also note that the next adsorber replacement should
be performed every three years.
Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068
and remove the adsorber from the Compressor.
Figure
5-1, connect the two self
in this section.
"Adding
5-3