HELIX 9600 Installation, Operation And Maintenance Instructions page 35

High voltage compressor
Table of Contents

Advertisement

Figure 3-2: 9600 Compressor Circuit Breaker Terminals (Cover Removed)
©2007 Helix Technology Corporation
See Detail A
Circuit Breaker
Terminal Cover
Rear Panel
NOTE: The phase order in which the conductor terminal lugs are
connected to circuit breaker terminals X, Y, and Z will be determined
during the Phase Check Procedure.
NOTE: For installation where one of the three phase legs is at or near
ground potential, connect that leg to terminal Y on the Compressor as
shown in
Figure
3-2.
10. Install the conductor terminal lugs to the circuit breaker terminals
X, Y, and Z as shown in
11. Torque the screws to 12 in.-lbs (0.14m-kg).
12. Allow enough cable to stay in the electrical enclosure to prevent
strain on the electrical connections and tighten the screws on the
cable strain relief.
13. Install the power source end of the power cable according to the
local electrical codes.
14. Install the circuit breaker terminal cover.
15. Proceed with Phase Check.
Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068