Multiple Cryo-Torr Cryopump Connections

Multiple Cryo-Torr Cryopump Connections
Make sure the Compressor power is OFF before making any
connections to the rear panel.

Helium Line Connections

The use of several compressors on a single manifold feeding a
common supply header and a common return header requires special
precautions. Contact CTI-CRYOGENICS for a review of the intended
installation and for specific technical instructions.
The use of a 9600 compressor on a manifold with other
CTI-CRYOGENICS compressor models requires a reduction of the
helium charge pressure to 200 - 210 psig charge pressure to avoid
helium safety valves from inadvertently venting. Refer to
Operation"

Power Cable Connections

3-14
Part Number 8040437, Revision 104, 06/12/2007, ECO Number 18068
CAUTION
CAUTION
and
"Section 5 - Maintenance"
1. Connect the Supply and Return lines to the 9600 Compressor as
described in
"Connecting/Disconnecting Helium Flex Lines"
this section.
2. Connect the Gas Return Line to the customer supplied helium
manifold and then to the Gas Return connector on the Cryo-Torr
Cryopump.
3. Connect the Gas Supply Line to the customer supplied helium
manifold and then to the Gas Supply connector on the Cryo-Torr
Cryopump.
1. Connect the Cryo-Torr power cable between the CRYOPUMP
ELECTRICAL OUTLET on the rear panel of the Compressor
and the CRYOPUMP ELECTRICAL INPUT on the Cryo-Torr
Interface as shown in
2. Connect the Cryo-Torr Power Cables between the CRYOPUMP 1,
"Section 4 -
for more information.
Figure 3-8
or
Figure
3-9.