Flow Connections To The Micro Vacuum Degasser - Agilent Technologies 1100 Series Reference Manual

Micro vacuum degasser
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Parts required
Preparation
1 Place solvent cabinet with the bottle(s) on top of the micro vacuum degasser.
2 Remove the front cover by pressing the snap fasteners on both sides.
Figure 5
3 If the micro vacuum degasser is not used with an Agilent 1100 Series pump,
4 Put the bottle head assemblies into solvent bottles containing your mobile
WAR NI N G
Installing the Micro Vacuum Degasser

Flow Connections to the Micro Vacuum Degasser

Flow Connections to the Micro Vacuum
Degasser
Solvent cabinet including solvent bottles (filled with solvent) and bottle head assemblies
Micro Vacuum degasser
Solvent outlet tubes
Syringe with adapter
Install the micro vacuum degasser
Removing the Front Cover
connect the waste tube from the accessory kit to the waste outlet and place
into your waste system.
phase.
When working on capillary or tube fittings solvents may leak out.
Please observe appropriate safety procedures (for example, goggles,
safety gloves and protective clothing) as described in the material
handling and safety data sheet supplied by the solvent vendor,
especially when toxic or hazardous solvents are used.
18