Input Accuracy - Omega DP606A User Manual

Universal 6/12 channel 1/4 din panel meter
Table of Contents

Advertisement

Section 8 - Input Accuracy
Measurement Ranges and Accuracies based on
Operating Temperature
Input Type
Description
Process Voltage
Process
Process Current
Type T/C
J Iron-Constantan
CHROMEGA™ -
Type T/C
ALOMEGA™
Type T/C
Copper-Constantan
CHROMEGA™
Type T/C
-Constantan
Type T/C
Pt/13%Rh-Pt
Type T/C
Pt/10%Rh-Pt
Type T/C
30%Rh-Pt/6%Rh-Pt
Type T/C
5%Re-W/26%Re-W
Type T/C
Nicrosil-Nisil
Pt, 0.00385, 100 Ω
2/3 Wire
Cu, 0.00427, 10 Ω
Wire
2
3
Ni, 0.00672, 120 Ω
Wire
2
3
* Absolute Maximum 3.3V (Process Voltage) or 30mA (Process Current).
Range
Accuracy (25°C)
0 to 1000 mV
± 1 mV
0 to 24.00 mA
± 10 µA
-150 to 0°C
± 1.0°C
0 to 1200°C
±1.0°C
-150 to 0°C
± 1.0°C
0 to -1372°C
± 1.0°C
-150 to 0°C
± 1.0°C
0 to 400°C
± 1.0°C
-150 to 0°C
± 1.0°C
0 to 1000°C
± 1.0°C
-50 to 0°C
± 1.0°C
0 to 1788°C
± 1.0°C
-50 to 0°C
± 1.0°C
0 to 1768°C
± 1.0°C
150 to 700°C
± 1.0°C
700 to 1820°C
± 1.0°C
0 to 2320°C
± 1.0°C
-150 to 0°C
± 1.0°C
0 to 1300°C
± 1.0°C
-200 to 850°C
± 1.0°C
-200 to 260°C
± 1.0°C
-80 to 260ºC
± 1.0ºC
Table 16 - Input Accuracy
Display Meters User's Guide
Operating Temperature
Accuracy (0 to 50°C)
Accuracy (-20 to 70°C)
± 1 mV
± 10 µA
± 2.0°C
± 1.0°C
± 2.0°C
± 1.0°C
± 2.0°C
± 1.0°C
± 2.0°C
± 1.0°C
± 2.0°C
± 1.0°C
± 2.0°C
± 1.0°C
± 2.0°C
± 1.0°C
± 1.0°C
± 2.0°C
± 1.0°C
± 1.0°C
± 1.0°C
± 1.0ºC
8
± 1 mV
± 10 µA
± 6.0°C
± 2.0°C
± 5.0°C
± 2.0°C
± 7.0°C
± 2.0°C
± 5.0°C
± 2.0°C
± 6.0°C
± 2.0°C
± 5.0°C
± 2.0°C
± 3.0°C
± 1.0°C
± 3.0°C
± 5.0°C
± 2.0°C
± 1.0°C
± 1.0°C
± 1.0ºC
30