Cleaning the Loss/Length Connectors

Always clean and inspect fiber connectors before making
connections. Use fiber optic solvent and optical-grade wipes
or swabs to clean connectors as follows:

Cleaning the Loss/Length OUTPUT Connector

1
Touch the tip of a fiber optic solvent pen or swab
soaked in solvent to a lint-free dry wipe or fiber
cleaning card.
2
Touch a new, dry swab to the solvent spot on the wipe
or card.
3
Push the swab into the connector, twist it around 3 to 5
times against the end-face, then remove and dispose of
the swab.
4
Dry the connector with a dry swab by twisting it
around in the connector 3 to 5 times.
5
Inspect the connector regularly with a fiber
microscope, such as the Fluke Networks FiberInspector
Video Microscope.
60
Cleaning the Loss/Length and Power Meter INPUT
Connector
1
Remove the connector adapter to expose the
photodiode lens (see Figure 24).
2
Use the method described in steps 1 and 2 under
"Cleaning the Loss/Length OUTPUT Connector" to
dampen a swab with fiber optic solvent.
3
Twist the damp swab around against the lens 3 to 5
times; then twist a dry swab around against the lens 3
to 5 times.