4-Wire Remote Or Remote Core Processor With Remote Transmitter Installations - Emerson Micro Motion 2700 Installation Manual

Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Installing the Transmitter continued
6. Rotate the transmitter clockwise so that the cap screws are in the locked position.
7. Tighten the cap screws, torquing to 20 to 30 in-lbs (2,3 to 3,4 N-m).
2.3.2

4-wire remote or remote core processor with remote transmitter installations

If you chose the 4-wire remote or the remote core processor with remote transmitter installation (see
Figure 2-1), see Figure 2-3 for a diagram of the mounting bracket supplied with the transmitter. Both
pipe mounting and wall mounting are shown. Ensure that the transmitter is mounted and oriented in a
way that will allow easy access to the terminals and to the display (if your transmitter has a display).
Figure 2-3
4-wire remote – Wall mount or pipe mount
Mounting bracket
(wall mount)
To mount the transmitter:
1. Identify the components shown in Figure 2-4. For dimensions, see Appendix A.
2. If desired, re-orient the transmitter on the bracket.
a. Remove the junction end-cap from the junction housing.
b. Loosen each of the four cap screws (4 mm) inside the junction housing.
c. Rotate the bracket so that the transmitter is oriented as desired.
d. Tighten the cap screws, torquing to 30 to 38 in-lbs (3 to 4 N-m).
e. Replace the junction end-cap.
3. Attach the mounting bracket to an instrument pole or wall. For pipe mount, two user-supplied
U-bolts are required. Contact Micro Motion to obtain a pipe-mount installation kit if required.
Transmitter Installation: Model 1700 and 2700 Transmitters
CAUTION
Twisting the core processor will damage the sensor.
To reduce the risk of damaging the sensor, do not allow
the core processor to rotate.
Mounting bracket
(pipe mount)
11

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Micro motion 1700