N is the entered pressure.
Two-point calibration is complete.

Save

Push the power button to save the calibration adjustment.
When you calibrate the Product, the results should typically be within
50 % of the total uncertainty. For calibration with the serial port, use
the VAL? command. A typical calibration is to be done at 20 %
increments first with ascending pressure and then with descending
pressures. The typical procedure for vacuum calibration is to be
done in this order:

User-Replaceable Parts

User-replaceable parts are shown in Table 4. For more information about these
items and their prices, contact a Fluke Calibration representative. See the
"Contact Fluke Calibration" section.
Gauge Cover
2700G-8004 DECAL BG100K 15PSI 0.1MPA
2700G-8005 DECAL BG200K 30PSI 0.2MPA
2700G-8006 DECAL BG700K 100PSI 0.7MPA
2700G-8007 DECAL BG2M 300PSI 2MPA
2700G-8008 DECAL BG3.5M 500PSI 3.5MPA
2700G-8009 DECAL BG7M 1000PSI 7MPA
2700G-8010 DECAL G20M 3000PSI 20MPA
2700G-8011 DECAL G35M 5000PSI 35MPA
2700G-8012 DECAL G70M 10000PSI 70MPA
POWER SUPPLY,SW,5W,85-264VAC,5 V@1 A,3 KV,W/MAINS
ADAPTERS,USB OUT,UNIV,WALL MOUNT
USB CABLE,USB TO RS232, 1.8M/5V WIRE END W/ CONN, 4
CONTACT, F CBL W/STRAIN RLF
2700G Manuals CD
AA Alkaline batteries, NEDA 15A IEC LR6
10
-6 psi, -12 psi,-3 psi and back to 0 psi.
Table 4. User-Replaceable Parts
Description
Note
Fluke Part Number
4201079
4201101
4201112
4201120
4201135
4201147
4201158
4201164
4201173
4201186
4252495
4258329
4150074
376756

Hide quick links: