A40 ............................................................. 2.5 mA - 5A
A40A ........................................................... 5A - 20A
Input Resistance ............................................. 91 Ω ± 1 %
Operating Input Voltage.................................. 250 mV to 500 mV
Maximum Non-Destructive Input .................... 50V rms

General Specifications

Warm-up Time ................................................ 30 minutes
Relative Humidity
Operating .................................................... 45 % to 50 °C 75 % to 45 °C 95 % to 30 °C
Storage ....................................................... <95 % non-condensing
Altitude
Operating .................................................... 3,050 meters (10,000 feet)
Non-Operating ............................................ 12,200 meters (40,000 feet)
Temperature
Operating .................................................... 0 °C to 50 °C
Calibration................................................... 15 °C to 35 °C
Storage ....................................................... 40 °C to 70 °C
EMI/RFI
Complies with ............................................. FCC Part 15 Subpart B, Class B; VDE 0871, Class B; ESD: EIA PN-
Surge .............................................................. ANSI C62.41-1980, Category A
Reliability ........................................................ MIL-T-2880D, paragraph 3.13.3
Size
Height ......................................................... 17.8 cm (7 in) standard rackmount + 1.5 cm (0.6 in)
Width........................................................... 43.2 cm (17 in)
Depth .......................................................... 63 cm (24.8 in)
Maximum Power Requirements
5790A ......................................................... 95 VA
With Wideband Option................................ 120 VA
Weight
5790A ......................................................... 24 kg (53 lb)
With Wideband ........................................... 24.5 kg (54 lb)
Line Power ...................................................... 47 Hz to 63 Hz; ± 10 % of selectable line voltages: 100 V, 110 V, 115
Safety.............................................................. Complies with UL1244 and IEC 348-1976 and IEC 1010 and CSA
Remote Interfaces .......................................... RS-232, IEEE-488
1361.
V, 120 V, 200 V, 220 V, 230 V, 240 V
C22.2 No. 231 and ANSI/ISA S82