Printing Calibration Reports - Fluke 5790A Operator's Manual

Ac measurement standard
Hide thumbs Also See for 5790A:
Table of Contents

Advertisement

6. This step needs to be done only if you are setting up the interface for remote
control of the 5790A. Press one of the softkeys under "Remote I/F" to set the
label to "TERMINAL" or "computer".
"Terminal" sets the remote port to expect a human operator to be using a terminal
attached to the RS-232 port to control the 5790A. This setting has the same effect
as specifying "TERM" in the SP_SET remote command parameter string.
"Computer" sets the remote port to expect a computer to be controlling the
5790A over the RS-232-C port. This setting has the same effect as specifying
"COMP" in the SP_SET remote command parameter string.
7. Press the "EOL" softkey to select the End of Line character.
8. To designate a character or character string as the EOF (End Of File) character,
press the "DONE Setting Port" and "Set EOF" softkeys.
9. Enter the decimal code of the ASCII (American Standard Code for Information
Interchange) character or characters to be designated as EOF. (Appendix C
contains a table of ASCII codes.) Verify that your selection is correct as shown
on the display.
10. To exit the setup menus, press the leftmost softkey four times.

7-4. Printing Calibration Reports

Calibration Reports show the shifts at various input levels and frequencies that are the
result of the most recent, or the previous calibration. How to print the reports and
examples of each type of report are provided next.
The calibration reports are test reports, not calculations of correction factors for you to
apply. Do not use the shifts printed on the reports as correction factors. After configuring
the serial interface for compatibility with your peripheral device and attaching an RS-232
null-modem cable, proceed as follows from the power-up state to transmit the report:
1. Press [UTIL MENUS], the "Cal" softkey, followed by the "Cal Reports" softkey. The
Control Display appears as follows:
2. Press one of the middle three softkeys to see the other report choice:
Note
Using the RS-232 Interface

Printing Calibration Reports

add55f.eps
7
7-5