Fluke 5790A Operator's Manual page 116

Ac measurement standard
Hide thumbs Also See for 5790A:
Table of Contents

Advertisement

5790A
Operators Manual
This mechanism (or *OPC? or *WAI) does not wait for a CAL_PR to
complete. Also see ISCR.
If you use this command for a non-measurement execution to complete, turn
off continuous triggering first. If you do not turn off measurements, *OPC
(or *OPC? or *WAI) will wait for a measurement to complete. Example:
"EXTRIG OFF; CAL_STORE ALL; *OPC".
Parameter
None
*OPC?
Returns a 1 after all pending operations are complete. This commands causes program
execution to pause until all operations are complete. (See also *WAI.)
Parameter
None
Response
(Integer) 1 after all operations are complete.
*OPT?
Query which hardware and software options are installed.
Parameter
None
Response
(IAD) WBND or 0
*PUD
(Protected user data command.) This command allows you to store a string of bytes in
nonvolatile memory. This command works only when the CALIBRATION STORE
switch is in the ENABLE position. Also see the RPT_STR command, which has a similar
function, except the RPT_STR string is printed on calibration shift reports.
Parameter
Binary Block Data
Example
To store the string "CAL LAB NUMBER 1" in the protected user data area.
6-10
*PUD #0CAL LAB NUMBER 1\<Line Feed with EOI\>
or
*PUD #216CAL LAB NUMBER 1
Note
Note