Linearity Calibration - Excell EX-2001 Racer Plus User Instructions

Table of Contents

Advertisement

EXCELL PRECISION CO., LTD.

5-2-4 Linearity Calibration

Before linearity calibration, make sure "Zero calibration" and "Weight Calibration" are properly
performed. Set the calibration switch to ON position.
C A L
Press
key
Choose
Press
key
Press
to next step
Press
to next step
Enter the correct weight value
Press
to next step
The calibrated weight value
is displayed when the
signal is stable
Press
to next step
◆ Refer to 5-4 Error Messages when any error message appears during the procedure.
EX-2001 Plus v1.3
. . . . . .
After finishing the first linearity calibration,
choose to continue or exit the linearity mode
by pressing
.
Five point linearity calibration.
Choosing from 1P ~ 5P by
pressing
calibration value at this point
value present at this point
Press
key to previous step
The weight value is displayed.
Press
key to previous step
15
key.
⇒ No linearity
⇒ linearity calibration
⇒ Increase the number of
the flashing digit
⇒ Decrease the number
of the flashing digit
⇒ Shift the flashing digit
one space to left
⇒ Shift the flashing digit
one space to right
⇒ Save the settings
⇒ Exit the settings