Honeywell XYR 6000 User Manual page 29

Onewireless wcx series wireless valve position sensor
Hide thumbs Also See for XYR 6000:
Table of Contents

Advertisement

Step
If not already open, remove cover lock clamp (If applicable) and cover per previous Table 1.
1
In Wireless Builder, set sensor's Write Lock to Unlocked.
2
In Wireless Builder, set the sensor's channel to OOS (Out of Service).
3
Use Authentication Device's Device Local Configuration buttons to navigate to the sensor's
4
CAL menu.
If the sensor is locked a LOCKED message will be displayed. Go to step 1.
If CAL menu is passcode protected, enter the passcode.
If the channel is not out of service a WRONG MODE message will be displayed. Go to step 2.
Select CAL CLR by scrolling through menu.
5
Press Enter to continue.
Display will say BUSY, then SUCCESS.
Exit the menu.
6
Reinstall cover and cable lock clamp (if applicable) per previous Table 1.
7
Revision 1
OneWireless XYR 6000 Wireless Valve Position Sensor User's Manual
1/15/09
Table 6 - Clear Calibration
Action
3. 2BPreparation
3.6. Calibrating the Sensor
19