4.2.3
Programming the Shaker
Press
or
The above example programmes the shaker to shake for 20 minutes after an initial delay of 10
minutes.
The display will revert back to the current temperature display following ~5 secs of no key activity
at any stage.
Once the Shaker is turned on by pressing the SHAKER ON/OFF button the red shaker LED will flash
indicating the shaker is operational. If a delay time has been programmed then the shaker will not
shake until this time has elapsed.
If the SHAKER TIMER button is pressed while the red shaker LED is flashing then the time remaining
in minutes will be displayed. If the shaker is stationary then the time displayed will be the remaining
delay time (d), if the shaker is shaking then the time displayed is the remaining shaker time (L).
Pressing the SHAKER ON/OFF button at any time during a programmed sequence will abort the
programme and restarting the shaker will start again from the delay time.
4.2.4
Continuous Shaking
The continuous shaking mode is entered by pressing the Shaker Timer button and pressing the  or
 keys until Cont is displayed. In this mode the shaker will continuously shake once turned on until
it is turned off again.
Agilent 1260 HT Sample Prep System Operation Manual
Display
Toggle between
mode sets using up
or down arrow
Shaker delay time
displayed in minutes.
Adjust using  and 
keys
Shaker time
displayed in minutes.
Adjust using  and 
keys
Comments