Problem:
Why does the monitor not display an SPO
changing the Masimo probe site?
Answer:
The Masimo probe is not recognized by the monitor.
Intervention:
When a Masimo sensor is repositioned at any time,
disconnect the cable from the sensor before
repositioning.
Reconnect the sensor to the interface cable after
proper patient preparation and placement.
SPO
2
Problem:
reading after
Why is the SPO
2
connecting the SPO
Answer:
No SPO
an unrecognized or defective probe.
Intervention:
Ensure the probe is attached to the interface cable
and the monitor.
Change the probe.
Change the cable.
If the issue remains, contact the hospital Biomedical
Engineering Department for service.