Relearn:
It is important to RELEARN anytime the
electrode configuration is adjusted.
Select the ECG parameter box.
Select RELEARN.
Remember to ensure that there is a clean ECG
signal displayed when relearning a rhythm.
Current ECG templates are now updated.
TROUBLESHOOTING and
FREQUENTLY ASKED QUESTIONS
ECG
If the Problem Continues:
Select the ECG Parameter Window.
Select VIEW ALL ECG to determine the best
display lead. Select the lead with the greatest
amplitude. (Remember that a minimum of
0.5 mV of amplitude in one direction, positive or
negative, is required for QRS detection).
Select DISPLAY LEAD: II.
Select the desired display lead.
Select LD ANALYSIS MULTI-LEAD.
Select SINGLE LEAD.
All arrhythmia interpretations will now be based
on this single ECG lead.