12 Lead Ecg Analysis - GE DASH 3000 V5 Quick Reference Manual

Bedside monitoring
Table of Contents

Advertisement

12 Lead ECG Analysis

NOTE: For the most accurate serial comparison, use the same
electrode configuration used on previous analysis.
To Obtain a 12-Lead on a
Patient:
Place patient in a supine
position.
Correctly identify and apply
all 10 electrodes.
Select the ECG Parameter
Window.
Select 12 LEAD ECG
ANALYSIS.
NOTE: Accurate demographics must be entered to ensure
proper analysis.
Select ADMIT INFO
Select CHANGE ADMIT
INFO.
Electrode Placement
ST
12 Lead ECG
Enter Information:
First Name: Completely
entered.
Last Name: Completely
entered.
Patient ID: Facility Specific,
(i.e., Medical Record
Number, Social Security
Number, etc.)
Sex
Front
12 Lead Electrode Placement
Leads
V1
Fourth intercostal space at the right
sternal border; right chest.
V2
Fourth intercostal space at the left
sternal border; left chest.
V3
Midway, between locations V2 and
V4; left chest.
V4
Mid-clavicular line in the fifth
intercostal space; left chest.
V5
Anterior axillary line on the same
horizontal level as V4; left chest.
V6
Mid-axillary line on the same
horizontal level as V4 & V5; left chest.
LA
Left deltoid or left wrist.
RA
Right deltoid or right wrist.
LL
Left thigh or left ankle.
LA
Right thigh or right ankle
Birth Date
Age: Correct age impacts
the analysis.
Height
Weight
Location ID: Identifies
origin of 12-lead to MUSE
for transmission.
(Set in defaults. For rover
modes, set in admit menu).
Site Number: Identifies
hospital to MUSE for
transmission.
(Set in defaults).
NOTE: Depending on facility policy, other fields may be
required; i.e., Technician ID, Optional Field, Order Number.
(continued on next card)
TAB
TRIM
COVER &
NON-TAB
TRIM
COVER &
NON-TAB
TRIM
TAB
TRIM

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dash 4000 v5

Table of Contents