Fig. 7-8
Positioning of the batteries into the battery tray is
completed with battery tray positioned in wheelchair
and partially pulled out. Illustrations are shown
without the wheelchair for clarification purposes
only.
60101831-A
Fig. 7-7
Fig. 7-9
1. Position the battery with battery connector bracket A
against the back edge of the battery tray B in the
orientation as shown.
Batteries MUST be installed in the orientation
shown so that the wiring harnesses can be
connected together.
2. Position the remaining battery between the battery on
the tray and battery stop C.
3. Connect batteries together D.
4. Connect battery straps E.
5. Slide the battery tray into the wheelchair
6. Install the rear shroud, front shroud and battery
retention bracket. Refer to 9.6 Removing/Installing the
Shrouds, page 70.
7. If Installing NEW batteries—The Battery Synchronization
Procedure MUST be performed within 24 hours of
powering on the wheelchair. Otherwise the battery
gauge will not display accurate battery charge levels.
The Battery Synchronization Procedure can be found in
the LiNX service manual and must be performed by a
provider or qualified technician.