Driving the unit

If you drive the unit, please observe the following:
❏ The power cable must not be connected to a wall outlet.
Driving with motor power, MOBILETT XP Hybrid only
The driving control has a function designed to prevent the MOBILETT from accel-
erating uncontrollably. This can occur in e.g. slopes or if the MOBILETT is forced
to go faster. The motor power will be shut-off until the MOBILETT slows down
again. This is to prevent possible accidents where the operator loses control of
the MOBILETT.
The maximum incline for the motordrive is 7
As long as one of the motor operation keys is pressed, the system blocks radia-
tion release.
Caution
Risk of injuring the operator!
The system drive may shut off suddenly, causing the unit to brake
◆ Be careful if you push the system manually while the motor is on.
◆ Check that the power cable is not connected to a wall outlet.
◆ Turn the mode selection switch to position battery operation.
– Main switch (key): On
°
.