Interval
12 months
12 months
12 months
12 months
12 months
12 months
Interval
12 months
24 months
24 months
48 months
12 months
MOBILETT
SPR8-230.621.01.03.02

Annual Maintenance

Maintenance every 12 months is necessary to keep the equipment safe and oper-
ational.
◆ We recommend that the maintenance is performed by the Siemens customer
service.
Check of the system
Operations to perform
Mechanical test
Functional check
Checking of image quality
Checking of kV/mAs accuracy
Electrical safety checks
USA only:
Checking of reproducibility
Check of system parts
Material
Collimator lamp
Battery pack (left) BK1/BK2
Battery pack (right) BK3/BK4
Cable winder complete
Batteries remote control (Option)
Functional and Safety Checks
Material No.:
8392016
08392024
08392032
06508746
1 x 9 V alkaline
73