Guidance and manu-
facturer's declaration -
electromagnetic
immunity
Immunity test
Electrostatic
discharge (ESD)
IEC 61 000-4-2
Electrical fast
transient/burst
IEC 61 000-4-4
Surge
IEC 61 000-4-5
Voltage dips, short
interruptions and volt-
age variations on power
supply input lines
IEC 61 000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61 000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
MOBILETT
SPR8-230.621.01.03.02
The Mobilett XP is intended for use in the electromagnetic environment specified
below. The customer or the user of the Mobilett XP should assure that it is used
in such an environment.
IEC 60601
Compliance
level
test level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input/
±1 kV for input/
output lines
output lines
±1 kV differential
±1 kV differential
mode
mode
±2 kV common
±2 kV common
mode
mode
<5% U
(>95%
n.a.
T
dip in U
) for 0.5
T
cycle
40% U
(60%
T
dip in U
) for 5
T
cycles
70% U
(30%
T
dip in U
) for 25
T
cycles
<5% U
(>95%
T
dip in U
) for 5
T
cycles
3 A/m
3 A/m
Technical Data
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the Mobilett XP
requires continued operation during power
mains inter-ruptions, it is recommended
that the Mobilett XP be powered from an
uninterruptible power supply or a battery.
The Mobilett XP has a rated input current of
more than 16 A per phase.
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital environ-
ment.
89