Excell EX-2005C User Manual page 25

Table of Contents

Advertisement

EXCELL PRECISION CO., LTD
3. COMMAND FORMAT C
Host
Command + Data
Slave
Write weight compared value
XXXXXX : value (6 bytes)
FW
Write target item of unload value
S2
Write SP2
UD
Write Under
ZB
Write Zero Band
PR
Write Peak value
Error messages
E1: Format command fault
E2: Setting parameters over range
E3: Command not recognised
Indicator poling address
If the indicator has an address configured in RS1 – 07 as 0, it will directly
respond to the command.
If the indicator has an address configured in RS1-07 as 1~99, it will only
respond to the messages prefixed the address.
For example: The indicator poling address is set to 02, it would send the weight
value only if it received the command:
EX-2005C
.
<CR>< LF>
Command + Data
□□
WS
S1
Write SP1
S3
Write SP3
LO
Write LO
HI
Write HI
OV
Write Over
@02RW<CR><LF>
24
<CR>< LF>
XXXXXX
□□
: Setting items
ZSME300000116