4
Configuring The Meter
4.3 USING READING CONFIGURATION ("RD.CF")
To select the calibration mode and units of measurement, follow these steps:
1.
Press the MENU button until the meter shows "RD.CF".
2.
Press the
• "R.1=2" For the two point buffer solution calibration (either 4 and 7, or 7 and 10
buffer solutions are used).
• "R.1=3" For three point buffer solution calibration (4, 7, or 7 and 10 buffer
solutions are used). (Default)
3.
Press the
4.
Press the
of the following:
• "R.2=C" The displayed temperature in degrees Celsius (Default)
• "R.2=F" The displayed temperature in degrees Fahrenheit.
5.
Press the
6.
Press the MENU button to store your choice. The meter momentarily shows
"STRD", followed by "CAL.2" or "CAL.3".
Refer to step 4.1, during this step a temperature compensation type was chosen.
If "OFF" was chosen then proceed to the "CAL.2" or "CAL.3" menu.
If "METR" was selected, connect the RTD sensor's leads to the RTD input
connector, and proceed to the "CAL.2" or "CAL.3" menu.
If "MANU" was selected, press the MENU button until the "TEMP" menu
displayed. (step 4.11 - Configuring Temperature).
22
T/mV button. The meter shows one of the following:
/pH button to scroll through available choices.
T/mV button to advance to the next submenu. The meter shows one
/pH button to scroll through available choices.