4.10 USING OUTPUT CONFIGURATION (OT.CF)

Use Output Configuration to select the following:
• To enable or disable the analog output
• To determine if the analog output is current or voltage

4.10.1 Enabling or Disabling the Analog Output

To enable or disable the analog output, follow these steps:
1.
Press the MENU button until the meter shows "OT.CF".
2.
Press the
T/mV button. The meter shows one of the following:
• "O.1=E" (Analog output enabled) (Default)
• "O.1=D" (Analog output disabled)
3.
Press the
/pH button to toggle between available choices.
4.
Press the
T/mV button to select the analog output as current/voltage or press the
MENU button to store your selection. The meter momentarily shows "STRD",
followed by "OT.S.O" (Output Scale and Offset) - refer to section 4.12.

4.10.2 Selecting Analog Output as Current or Voltage

1.
Press the
T/mV button. The meter shows one of the following:
• "O.2=C" (Analog output = current) (Default)
• "O.2=V" (Analog output = voltage)
2.
Press the
/pH button to toggle between available choices.
3.
Press the
T/mV button to go back to selecting analog output as enabled or
disabled or press the MENU button to store your choices. The meter momentarily
shows "STRD", followed by "OT.S.O" (Output Scale and Offset) - refer to section
4.12.
Configuring The Meter
4
29