3.4 PANEL MOUNTING

FRONT BEZEL
1. Cut a hole in your panel, as
shown in Figure 3-4. For
specific dimensions refer to
Figure 3-5.
2. Insert the meter into the
hole. Be sure the front bezel
gasket is flush to the panel.
3. Slide on mounting bracket to
secure.
4. Proceed to Section 3.5 to
connect your sensor input
and main power.
CONNECTOR
LABEL
PRODUCT
LABEL
GASKET
Figure 3-4. Meter - Exploded View
Getting Started
CASE
1,5
R(.06)
45,00 + 0,61/-0,00
(1.772 + .024/–.000)
4 PLCS
92,00 + 0,81/–0,00
(3.622 + .032/–.000)
NOTE: Dimensions in Millimeters (Inches)
Figure Figure 3-5. Panel Cut-Out
PANEL THICKNESS
6,4 (.25) MAX
0,8 (.03) MIN
3
15