Setting the Number of Points

The number of points is the number of data items collected in one sweep.
It can be set to any number from 2 to 1601 for each channel
independently.
To obtain a higher trace resolution against the stimulus value,
choose a larger value for number of points.
To obtain higher throughput, keep the number of points to a smaller
value within an allowable trace resolution.
To obtain higher measurement accuracy after calibration, perform
calibration using the same number of points as in actual
measurements.
When the upper limit of the number of channels and traces is
set to 1Ch / 4 Tr 20001 Points, the maximum number of
measurement points will be 20001.
Setting the number of points
1. Press
Channel Next/Channel Prev
2. Press
Sweep Setup key.
3. Click Points, then input the desired number of points.

Setting the Sweep Time

Sweep time is the time it takes to complete a sweep for each stimulus
(source) port. Two modes are available for setting the sweep time: manual
sweep time mode and automatic sweep time mode.
Manual
Sweep Time
Mode
Automatic
Sweep Time
Mode
92
key, then click Power > Slope [xxx dB/GHz] where
ENTRY
In this mode, the sweep time is set manually. Once the
sweep time is set, changes in measurement conditions will
not affect the sweep time as long as it is within the analyzer's
capability. If the sweep time becomes lower than the
analyzer's lower sweep time limit, the sweep time will be
reset to the shortest time within the conditions. If the sweep
time exceeds the analyzer's upper sweep time limit, the
sweep time will be reset to the longest time within the
conditions.
The sweep time is always kept to the shortest time possible
with the current measurement conditions.
to select the channel for which you
block keys on the front panel.
keys to select the desired channel.