Power-Up Configuration

The Tester has two types of configurations: standard and power-up. The
standard configuration, which is programmed into the Tester and can always
be retrieved, includes the following settings:
FCN
V
Volts Function selected.
DISP
WAVE Waveform Screen Mode selected.
THD
%RMS
Total Harmonic Distortion computed as percentage of
the total RMS value.
CLAMP
80I-500S
OTHER specifies any current probe other than the 80i-500s. The Tester
has been calibrated to provide either compensation for the 80i-500s or a
flat response for other probes.
P REF V Voltage Phase Reference selected.
For Model 41B, KBAUD 9.6 is the standard configuration for Print or Send,
and EPSON is the standard configuration. The Tester always selects
autoranging at power-up.
The power-up configuration, which is initially the same as the standard
configuration, can be changed by making entries from the configuration
screen. The Tester sets up to the power-up configuration each time it is turned
on. You can change the power-up configuration using the following
E
procedure:
1. Press
while pressing
<
>
screen appears. (See Figure 17.)
E
2. Press
and
to select the currently highlighted item and proceed to the next
line.
The 80i-500s Current Probe is selected.
O
to turn the Tester on. The configuration
to highlight different items on the same line. Press
Power-Up Configuration
37