Fluke 434-II User Manual page 169

Three phase energy and power quality analyzer
Hide thumbs Also See for 434-II:
Table of Contents

Advertisement

Introduction
Model Numbers:
Fluke 434-II, Fluke 435-II, Fluke 437-II
Short Description:
3-Phase Energy & Power Quality Analyzer
Memory.
Fluke 43x-II has the following memory devices:
1. RAM 8M x 16, U901, type: e.g. MT47H64M16HR-25IT:H,
2. Video RAM 256k x 16, D1001, type: e.g. CY62146EV30LL,
3. Flash-ROM 16M x 2, U1100, U1101, type: e.g. MW29W160EB,
4. FIFO (First In First Out) RAM 2kB, U801, type: e.g. SN74V235-7PAG,
5. SRAM 16 Mb x 2, D1100, D1103, type: e.g. CY62167DV30LL,
6. SD Memory Card. Contains: all datasets, screens, and logging data.
info@Fluke-Direct.ca
Instrument Security Procedures
contains: temp storage of measuring data
contains: storage of data to be displayed on LCD-screen.
contains: the instrument's embedded software and calibration data.
Also Analyzer settings such as Config, Freq, Vnom, Limits, and Current Clamp data
that differ from Factory Default are stored here.
contains: data to be exchanged between DSP and Microcontroller.
contains: temporary data storage for microcontroller.