Control Drawing Of Sc202S Ff/Pb Specification (Csa) - YOKOGAWA SC202G User Manual

2-wire conductivity or resistivity transmitter
Hide thumbs Also See for SC202G:
Table of Contents

Advertisement

2-15. Control Drawing of SC202S FF/PB Specification (CSA)

+
Safe area
I.S.
Apparatus
interface
-
Safe area
Sensor(s) are a thermocouple, RTD's, passive resistive switch devices, or is CSA entity
approved and meet connection requirements.
Electrical data of the SC202S-F & SC202S-P:
Supply and output circuit:
-
Maximum input voltage Ui = 24 V
Maximum input current Ii = 250 mA
Maximum input power Pi = 1.2 W
Effective internal capacitance Ci = 220 pF; Effective internal inductance Li = 0 μH.
or
FISCO field device
Maximum input voltage Ui = 17.5 V
Maximum input current Ii = 380 mA
Maximum input power Pi = 5.32 W
Effective internal capacitance Ci = 220 pF;
Effective internal inductance Li = 0 μH.
Sensor input circuit:
-
Maximum output voltage Uo = 14.4 V; Maximum output current Io = 13 mA
Maximum allowed external capacitance Co = 59 nF
Maximum allowed external inductance Lo = 200 mH
Any CSA approved I.S. interface may be used that meets the following requirements:
Uo
24 V
Io
250 mA
Po
1.2 W
Co
220 pF + Ccable; Lo
or
FISCO field device
Uo
17.5 V
Io
380 mA
Po
5.32 W
Co
220 pF + Ccable; Lo
Installation should be in accordance with Canadian Electrical Code, Part I or CEC, Part I.
Maximum safe area voltage should not exceed 250 Vrms.
Electrical data of the SC202S-B & SC202S-D (non-incendive):
For Class I, Div.2, Group ABCD the CSA approved I.S. interface is not required, and the
sensor input circuit is non-incendive having the parameters:
Maximum output voltage Uo = 14.4 V; Maximum output current Io = 13 mA
Maximum allowed external capacitance Co = 2.9 μF
Maximum allowed external inductance Lo = 450 mH
CSA Ex ia Class I, DIV. 1, Group ABCD
T4 for ambient temp.
Ui = 24 V
Ii = 250 mA
Pi = 1,2 W
SC202S-F
or SC202S-P
+
I.S.
certified
Terminator
0 μH + Lcable
0 μH + Lcable
55 C
or
Ui = 17,5 V
Ii = 380 mA
Pi = 5,32 W
-
-
+
Transmitter
Transmitter
Zone 0 or 1
Hazardous area
Specifications 2-19
Sensor
Connections
I.S.
certified
Terminator
-
+
IM 12D08B02-01E