5-18 Parameter setting

5-7. Communication setup

Code 60
*COMM.
mA
*ADDR.
mA
Code 61
*HOUR
*MINUT
*SECND
*YEAR
*MONTH
*DAY
Code 62
*ERASE

5-8. General

Code 70
*LOAD

5-9. Test and setup mode

Code 80
*TEST
NOTE: Attempting to change data in service code, 80 and above without the proper instructions and
equipment, can result in corruption of the instrument setup, and will impair the performance of
the unit.
IM 12D08B02-01E
The settings should be adjusted to suit the communicating device connected to
the output. The communication can be set to HART® or to PH201*B distributor
(see Appendix 2).
Select address 00 for point to point communication with 4-20mA transmission.
Addresses 01 to 15 are used in multi-drop configuration (fixed 4mA output).
The clock/calendar for the logbook is set for current date and time as reference.
Erase logbook function to clear the recorded data for a fresh start. This may
be desirable when re-commissioning an instrument that has been out of service
for a while.
The load defaults code allows the instrument to be returned to the default set
up with a single operation. This can be useful when wanting to change from
one application to another.
Not used.