Calibration With Hold Active - YOKOGAWA SC202G User Manual

2-wire conductivity or resistivity transmitter
Hide thumbs Also See for SC202G:
Table of Contents

Advertisement

6-3. Calibration with HOLD active

YES
NO
MODE
HOLD
YES
NO
YES
HOLD
YES
NO
Put the sensor in standard
solution. Press YES.
After the indication is stable,
using the >, , ENT key.
MODE
MEASURE
CAL
DISPLAY
HOLD
MODE
ENT
HOLD
set the value
Select the flashing digit with the > key.
Increase its value by pressing the key
When the correct value is displayed,
press ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press YES.
HOLD will be displayed. Press NO to turn off
HOLD and return to the measuring mode.
HOLD
ENT
Calibration 6-3
Press the MODE key.
The legend CALIB
appears, and the YES/NO
key prompt flags flash.
HOLD
ENT
HOLD
YES
HOLD
YES
IM 12D08B02-01E
ENT
NO
NO