10.10 The EDM beam axis

The distance measurement (EDM) beam axis is adjusted to be aligned to the sighting axis
of the telescope, but it can be changed a little in case of rapid temperature change,
shock or aging. Check your instrument by following procedures.
[Checks]
Install the instrument on the tripod and level it at the distance of approx. 50m from
1
the wall.
Displace the target plate attached to the end of this manual. Place the target plate
2
adjusting its center to the center of telescope cross line and to be about horizontal to
the instrument.
Turn the power on, and confirm the [TARGET] is set to the reflecting sheet mode
3
("S" will be indicated at the top of display, left side of the battery remains sign when it
is that.) When it is not, press [F2] [TARGET] and set to eflecting sheet mode (Reference
p39 - 6.1 Target setting).
Press [F1] [MEAS], then the laser spot appears on the target plate. If the "Center" of the
4
laser spot is within the internal circle (10mm) of the target plate at this moment,
the adjustment is unnecessary.
• The laser spot disappears in approx. 20 seconds after pressing [F1][MEAS].
Press [MEAS] again, if it is necessary.
20mm
[Example: Adjustment is necessary]
[Adjustments]
At the procedure 4. above, if the "Center" of laser spot is not within the internal circle
(10mm) of the target plate, the adjustment is necessary. Please contact your PENTAX
dealer.
66
10mm
Laser spot "center"
Target Plate
approx. 50m
Internal circle
20mm
[Example: Adjustment is unnecessary]
10mm