9.8.2.1 Baud Rate

The baud rate is the first parameter in the menu. The baud rate setting
determines the serial communications transmission rate.
The baud rate parameter is indicated by,
bAUd
2400 b
The baud rate of the serial communications may be programmed to 300, 600,
1200,2400, 4800, or 9600 baud. Use "UP" or "DOWN" to change the baud rate
value.
4800 b
Press "SET" to set the baud rate to the new value or "EXIT" to abort the
operation and skip to the next parameter in the menu.

9.8.2.2 Sample Period

The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature
measurements transmitted from the serial interface. If the sample rate is set to 5,
the instrument transmits the current measurement over the serial interface
approximately every five seconds. The automatic sampling is disabled with a
sample period of 0. The sample period is indicated by,
SPer
SP= 1
Adjust the value with "UP" or "DOWN" and then use "SET" to store the sample
rate to the displayed value. "EXIT" does not store the new value.
SP= 60

9.8.2.3 Duplex Mode

The next parameter is the duplex mode. The duplex mode may be set to full
duplex or half duplex. With full duplex any commands received by the calibrator
via the serial interface will be immediately echoed or transmitted back to the
device of origin. With half duplex the commands will be executed but not echoed.
The duplex mode parameter is indicated by,
dUPL
9-10
Find Quality Products Online at:
Flashes "bAUd" and then displays the setting
Current baud rate
New baud rate
Flashes "SPEr" and then displays the setting
Current sample period (seconds)
New sample period
Flashes "dUPL" and then displays the setting