Baud Rate - Fluke 7312 User Manual

Triple point of water maintenance bath
Table of Contents

Advertisement

7312 TPW Maintenance Bath
User's Guide
9.12
9.12.1
52
tP CO
Triple point of water cutout parameter
Press the "SET" button to access the activation selection. Press the "UP" or
"DOWN" button to turn the cutout on or off and then press the "SET" button.
Normally the cutout is ON.
Press the "EXIT" button to return to the operating parameters menu.
Triple point of water set-point selection parameter
tP SP
Press the "SET" button to access the parameter setting.
C -0.30
Cutout temperature in degrees C or F
To change the cutout temperature, use the "UP" or "DOWN" button. To accept
the new cutout set-point, press the "SET" button.
Press the "EXIT" button to return to the operating parameters menu.
Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SErIAL
Serial RS-232 interface parameters menu
The serial interface parameters menu contains parameters, which determine the
operation of the serial interface. The parameters in the menu are – BAUD rate,
sample period, duplex mode, and linefeed.

BAUD Rate

The BAUD rate is the first parameter in the menu. The BAUD rate setting de-
termines the serial communications transmission rate.
The BAUD rate parameter is indicated by,
Serial BAUD rate parameter
BAUd
Press the "SET" button to choose to set the BAUD rate. The current BAUD rate
value is displayed.
1200 b
Current BAUD rate
The BAUD rate of the bath serial communications may be programmed to 300,
600, 1200, or 2400 BAUD. Use the "UP" or "DOWN" button to change the
BAUD rate value.
New BAUD rate
2400 b

Advertisement

Table of Contents
loading

Table of Contents