7381
User's Guide
4. Push UP or DOWN to change the value.
HGb Po
5. Push SET to access the parameter setting
St
=
170
6. Push UP or DOWN to change the value (if in On mode).
7. Push SET to store the value or push EXIT to discard the changes and escape to the menu.

Serial Interface Parameters

Push UP to access the Serial Interface Parameters menu from the Operation Parameters menu.
SErIAL
indicates the serial RS-232 interface parameters menu.
The serial interface parameters menu contains parameters that determine serial interface
operation:

BAUD rate

sample period
duplex mode
linefeed
BAUD Rate
BAUD rate (
bAUd
) is the first parameter in the menu. The BAUD rate setting determines the serial
communications transmission rate. Program the Bath BAUD rate serial communications to 300,
600, 1200, or 2400 BAUD.
1. Push SET to choose to set the baud rate. The current BAUD rate value shows.
1200 b
2. Push UP or DOWN to change the BAUD rate value.
3. Push SET to store the new baud rate value, or EXIT to abort the operation and go to the next
menu parameter.
38
HGB setting parameter indicator
Current HGb setting value
Current BAUD rate