Cut-Out Reset Mode; Serial Interface Parameters; Baud Rate - Fluke 7008 User Manual

Calibration bath
Table of Contents

Advertisement

7008
Users Guide
Press "SET" to enter the menu. The operating parameters menu contains the
cut-out reset mode setting.

9.11.1 Cut-out Reset Mode

The cut-out reset mode determines whether the cut-out resets automatically
when the bath temperature drops to a safe value or must be manually reset by
the operator.
The parameter is indicated by,
CtorSt
Press "SET" to access the parameter setting. Normally the cut-out is set for
manual mode.
Cto=rST
To change to automatic reset mode press "UP" and then "SET".
Cto=Auto

9.12 Serial Interface Parameters

The serial RS-232 interface parameters menu is indicated by,
SErIAL
The Serial interface parameters menu contains parameters which determine the
operation of the serial interface. These controls only apply to baths fitted with the
serial interface. The parameters in the menu are — BAUD rate, sample period,
duplex mode, and linefeed.

9.12.1 Baud Rate

The baud rate is the first parameter in the menu. The baud rate setting
determines the serial communications transmission rate.
The baud rate parameter is indicated by,
BAUd
Press "SET" to choose to set the baud rate. The current baud rate value will then
be displayed.
1200 b
9-12
Cut-out reset mode parameter
Cut-out set for manual reset
Cut-out set for automatic reset
Serial RS-232 interface parameters menu
Serial baud rate parameter
Current baud rate