Specifications And Environmental Conditions; Specifications - Fluke 7102 User Manual

Micro-bath
Hide thumbs Also See for 7102:
Table of Contents

Advertisement

3.1 Specifications

Range
Accuracy
Stability
Uniformity
Resolution
Operating Temperature
Heating Time
Cooling Time
Well Size
Exterior Dimension
Weight
Power
Readout
Controller
Thermal Electric Devices (TED)
Cooling
Fault Protection
Safety
Fuse Rating
Find Quality Products Online at:
Specifications and Environmental
Table 2. Specifications
GlobalTestSupply
www.
Conditions
−5 to 125 °C (23 to 257 °F)
±0.25 °C
±0.015 °C at −5 °C (oil, 5010)
±0.03 °C at 121 °C (oil, 5010)
±0.02 °C
0.01 °C/F
5 to 45 °C (41 to 113 °F)
25 °C to 100 °C (77 °F to 212 °F): 30 minutes
25 °C to 0 °C (77 °F to 32 °F): 30 minutes
2.5" dia. x 5.5" deep (64 x 139 mm)
(access opening is 1.9" [48 mm] in diameter )
12" H x 7.2" W x 9.5" D (31 cm x 18 cm x 24 cm)
15 lb. (6.8 kg) with fluid
115 VAC ( ±10 %), 1.8 A, or 230 VAC ( ±10 %), 0.9
A, switchable, 50/60 Hz, 200 W
Switchable °C or °F
Digital controller with data retention
150 W
Fan and Thermal Electric Devices (TED)
Sensor burnout and short protection
Overvoltage (Installation) Category II, Pollution
Degree 2 per IEC 61010-01
115 V: 250 V 3A SB (slow blow)
230 V: 250 V 1.6 A T (time delay)
.com