Specifications And Environmental Conditions; Specifications; Environmental Conditions - Fluke 6050H User Manual

Calibration bath
Hide thumbs Also See for 6050H:
Table of Contents

Advertisement

3
Specifications and Environmental
Conditions
3.1

Specifications

Range
Stability
Uniformity
Temperature Setting
Set-Point Resolution
Display Temperature
Resolution
Digital Setting Accuracy
Digital Setting
Repeatability
Heaters
Access Opening
Depth
Wetted Parts
Power
Volume
Weight
Size
Automation Package
Safety
3.2

Environmental Conditions

Although the instrument has been designed for optimum durability and trou-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean-
ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
• ambient temperature range: 5 - 50°C (41 - 122°F)
• ambient relative humidity: maximum 80% for temperature <31°C, de-
creasing linearly to 50% at 40°C

3 Specifications and Environmental Conditions

180°C to 550°C
±0.002°C at 200°C (salt)
±0.004°C at 300°C (salt)
±0.008°C at 550°C (salt)
±0.005°C at 200°C (salt)
±0.020°C at 550°C (salt)
Digital display with push-button data entry
0.01°C; high-resolution mode, 0.00018°C
0.01°C
±1°C
±0.02°C
400/1200/2000 Watts
5" x 10" (127 x 254 mm)
12" (305 mm)
304 stainless steel
230 VAC (±10%), 50/60 Hz, 10 A, 2200 W
7.1 gallons (27 liters), requires 112 lb. of bath salt
180 lb. (82 kg)
28.5" H x 20.4" W x 24.5" D
(724 x 518 x 622 mm)
Interface -it software and RS-232 computer interface are available for set-
ting bath temperature via remote computer. For IEEE-488, add the
2001-IEEE to the automation package.
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per
IEC61010-1
Specifications