Or Shoes - Miele PG 8527 Operating Instructions Manual

Washer-disinfector
Table of Contents

Advertisement

Application

OR shoes

OR shoes made of
temperature-sensitive material and/or
insoles must be thermochemically
cleaned and disinfected at 60°C
(program CHEM-DISIN).
A thermal disinfection program
(SHOE-TD-75/2) can be used as long
as the manufacturer has confirmed the
thermal stability of the product(s).
For information on the efficacy of
thermochemical disinfection, contact
the disinfectant manufacturer directly.
OR shoes should only be cleaned
and disinfected in a
washer-disinfector that has been
installed for that application.
If OR shoes are processed with
other applications in a
washer-disinfector, the operator
must assess the risks associated
with this.
34
Mobile unit E 975/1 with a suitable
insert can be used for this purpose,
e.g.
– Insert E 930 for up to shoe size 43
(European), size 10 (US) can be
placed in the lower level of mobile
unit E 975/1,
– Insert E 931 for up to shoe size 48
(European), size 15 (US).
When OR shoes are cleaned, large
quantities of lint can accumulate. For
this reason, be sure to check the
wash cabinet filters frequently and
clean as needed (see the "Cleaning
the wash cabinet filters" section in
"Maintenance of the
washer-disinfector").